User login page. Forgot Password.
- What is the association between acute phase reactants and COVID-19? Check out Med Lab Study Hall’s latest blog to find out!
- Downloading MagicBox Study Free Thank you for using our Mac software library. Unfortunately, there is no direct download for the Mac version of MagicBox Study. To download the product, proceed to the App Store via the link below.
The Team AMU decided to study 5 different promoters from the collection in table 1 (J23117, J23106, J23118, J23101 and J23100) and compare also the T7 promoter Bba_K525998 that we used for some of our constructions (J23117, J23106, J23118, J23101 and J23100). We could then measure the strength of the promoter we used under standard conditions in a well characterized assay.Each promoter was integrated into a construction expressing GFP Bba_I13504 with a strong/weak RBS (BBa_B0034).
Measurement methods
- - TECAN microplate reader
- - Fluorometer (SPEX fluorlog)
- - Magicbox (home made filter based fluorimeter)
- - Confocal Microscope
Each machine (if possible) was calibrated using a standard solution of the dye Atto488. This allow us to compare the measurements on the different instruments. Unfortunately the confocal microscopy couldn’t be calibrated this way as with the solution there was nothing to see, no signal and impossible to focus the microscope on nothing!!
The first figure shows the calibration curve, diluting the Dye Atto488 in either buffer (PBS) or culture media (LB). In each case we have a good linear relationship between dye concentration and measured fluorescence. We observe a small but significant difference in the slope of the standard curve depending on the buffer showing a slight reduction of fluorescence in LB buffer.The results obtained for the different promoters are shown here. In agreement with the previous measurements for RFP the J23100 promoter is the strongest and the J23117 the weakest of those in the promoter collection. Our T7 shows very low basal expression barely different from the negative control (477+/-4 vs 224+/-316 for the control) this means that in the absence of induction in the correct strain there is probably no leakage and the promoter does not work at all.
Fluorometer results
The fluorimeter we used is much more sensitive than the microplate reader. So the samples used, both for the standard curve went only to 0.0175 ug/ml as opposed to 0.7 ug/ml for the microplate reader. For the cell measurements samples were also 10 times more dilute. As in the microplate reader, and can be seen in the figure, the standard curve with Atto488 is linear, with very little dispersion of the points. This curve shows, that under our measurement conditions the fluorimeter is about 100 times more sensitive than the microplate reader, as measured by the slope ; and 5 times more sensitive if we consider the concentration necessary to obtain twice the background measurement (0.5 vs 2.2 ng/ml).
Below we show the results obtained for the different promoters measuring fluorescence from diluted cell suspensions. These results are completely in agreement with those obtained using the microplate reader, and from the measurements of RFP. In view of the lower dispersion of the standard curve, and the greater sensitivity of the instrument we expect these results to be slightly better than those obtained using the microplate reader.
The microplate reader and fluorimeter are commercial instruments designed to measure fluorescence in solution. We wished to test a prototype instrument that we constructed for measuring GFP. The instrument and its use are described in the methods and protocols section.The standard curve with the Dye shows there is a reasonable linear relationship between the measured fluorescence and the dye concentration (R²=0.96) though this is much less good than we obtained with the fluorimeter or the microplate reader suggesting the measurements are less accurate and reproducible. The instrument is also less sensitive than the other machines with a measurement of twice the background at 12.4 ng/ml.
The measurements of cell suspensions shown here are less obviously consistent with the RFP measurements than obtained the machines used above. Various possibilities could be suggested for these less consistent results. First, as observed with the standard curve, a certain dispersion in the measurements probably arising from an imperfectly optimized protocol. Second, there is also a certain amount of stray excitation light detected by the apparatus and this may also degrade the performance. However the results obtained with this simple and cheap device could probably be improved with a little more optimization.
Confocal microscopy results
A microscope allows us to observe individual cells, this can give further information on the bacteria containing the different expression systems. Are all the cells expressing the same level of protein or are some bacteria much more fluorescent than others ? Is the protein expressed as a soluble cytoplasmic form or accumulating in inclusion bodies ? Is the physiology of the bacteria seriously modified resulting in bacteria with a strange appearance.
In the figure here we show some bacteria expressing GFP from the construction with the strong (1791AU in the table) J23101 promoter. While most of the bacteria appear normal in shape and have a homogeneous green colour (when they are well fixed to the coverslip) some are longer and less green (botom of image). This means that the protein is expressed in the cytoplasm, and so does not accumulate in inclusion bodies, but does cause some perturbations to the bacterial physiology resulting in fiamentation. This change in the shape of bacteria was only observed with the strongest promoters.
In the figure here we show some bacteria expressing GFP from the construction with the strong (1791AU in the table) J23101 promoter. While most of the bacteria appear normal in shape and have a homogeneous green colour (when they are well fixed to the coverslip) some are longer and less green (botom of image). This means that the protein is expressed in the cytoplasm, and so does not accumulate in inclusion bodies, but does cause some perturbations to the bacterial physiology resulting in fiamentation. This change in the shape of bacteria was only observed with the strongest promoters.
In order to quantify the fluorescence we selected bacteria in the image and measured the integrated fluorescence intensity for the different bacteria, correcting for changes in the laser intensity used to avoid saturation of the detector. The results from this protocol are shown below and show that there is a less good correlation than with the RFP measurements and those from the microplate reader and the fluorimeter. This lack of correlation could come from various experimental factors, consistency of focus, accurate estimates of cell number thrown off by changes in cell shape, bad statistics (not measureing enough cells).
Magicbox Study Music
Different promoters and expression levels.
First we find that the T7 promoter we used has no leakage and no detectable GFP is synthesised.
Comparing the different fluorescence levels qualitatively the results are essentially consistent with those obtained using RFP though the quality (accuracy and precision) of the measurements is better for the TECAN microplate reader and the fluorimeter.
TECAN | Fluorometer | Magicbox | Confocal Microscopy | |
---|---|---|---|---|
J23100 | +++++ | +++++ | +++++ | +++++ |
J23101 | ++++ | ++++ | +++ | +++++ |
J23118 | ++++ | +++ | +++++ | +++++ |
J23106 | ++ | ++ | + | + |
J23117 | + | + | + | + |
P T7 | - | - | - | - |
To quantitatively compare the measurements with GFP expression and those previously obtained with RFP we have plotted one against the other for our 2 best protocols. If the different promoters have the same strength in the different contexts we would expect a linear relationship. While the graph shows an approximately linear relationship it is imperfect, first the scatter suggests a small effect of the coding gene on the strength of the promoter, second the levelling of in the TECAN microplate measurements suggests an instrumentation problem (saturation of some sort) associated with our experimental conditions.
Magicbox Studying
Different machines
Magicbox Study Island
We obtained fluorescence measurements with 4 different machines. The microplate reader and fluorimeter appear to give the best and easiest to interpret results with the possibility of using a calibration curve to obtain absolute concentrations of GFP. This would need a standard GFP solution for the calibration measurements rather than Atto488.
The Magicbox measurements were less accurate and precise. However it may be possible to improve the quality of these measurement by reducing the stray light detected and optimizing the quantification protocol. With more programming skills a control of exposure time would also probably improve performance. Offering a cheap measurement system for GFP.
The microscope measurements did not give particularly good quantitative measurements. However they did provide useful additional information on cell to cell variability, sub-cellular localization of GFP and changes in bacterial morphology.
The Magicbox measurements were less accurate and precise. However it may be possible to improve the quality of these measurement by reducing the stray light detected and optimizing the quantification protocol. With more programming skills a control of exposure time would also probably improve performance. Offering a cheap measurement system for GFP.
The microscope measurements did not give particularly good quantitative measurements. However they did provide useful additional information on cell to cell variability, sub-cellular localization of GFP and changes in bacterial morphology.